Respected readers, authors and reviewers, you can add comments to this page on any questions about the contribution, review, editing and publication of this journal. We will give you an answer as soon as possible. Thank you for your support!
Zhejiang provincial Institute for Drug Control, Hangzhou 310004
Received Date: 2003-10-28
Abstract
OBJECTIVE An HPLC method was developed for determination of dl-Tetrahydropamatine in Yan-husuo. METHOD:The column was ODS Agilent zobax Extend C18(4.6×250mm,5μm),mobile phase was Meth-anol-0.1%. phosphoric acid(adjust pH 6.0 with Triethylamine)(55:45). Detection wavelength was set at 280nm. RESULT:Linearity range was 115-1376ng. The average recovery was 99.2%, RSD=1.5%(n=6). CONCLUSION The method is accurate and can be used for quality control of Yanhusuo.
Abstract: OBJECTIVE An HPLC method was developed for determination of dl-Tetrahydropamatine in Yan-husuo. METHOD:The column was ODS Agilent zobax Extend C18(4.6×250mm,5μm),mobile phase was Meth-anol-0.1%. phosphoric acid(adjust pH 6.0 with Triethylamine)(55:45). Detection wavelength was set at 280nm. RESULT:Linearity range was 115-1376ng. The average recovery was 99.2%, RSD=1.5%(n=6). CONCLUSION The method is accurate and can be used for quality control of Yanhusuo.
MA Lin-ke. Determination of dl-Tetrahydropamatine in Yanhusuo by HPLC[J]. Journal of Pharmaceutical Practice and Service, 2003, (6): 351-353.
Citation:
MA Lin-ke. Determination of dl-Tetrahydropamatine in Yanhusuo by HPLC[J]. Journal of Pharmaceutical Practice and Service, 2003, (6): 351-353.